美国纺织品面料标准 ASTM F 2100 医用面罩用材料性能规格测试
| 更新时间 2024-11-28 20:00:00 价格 请来电询价 测试周期 5-7天 寄样地址 深圳宝安 价格费用 电话详谈 联系电话 17324413130 联系手机 17324413130 联系人 罗卓文 立即询价 |
Test item 测试名称: 医用面罩用材料性能规格 Mask Performance Rating
Test info 测试内容:
Bacterial Filtration Efficiency (BFE) only 细菌过滤效率 , EN 14683 Annex B, ASTM F 2101, 15pcs (无菌样品) |
Differential pressure only , Delta P , 气压差, ASTM F 2100, 10pcs |
Particle Filtration Efficiency,PFE: Latex particle challenge 乳胶颗粒过滤效率, ASTM F 2299, 15pcs |
ASTM F 1862 Synthetic blood fluid penetration resistance for face masks 血液穿透阻力+ 防溅阻力, ASTM F 1862, 40pcs |
16 CFR Part 1610 Flammability test 阻燃, 20pcs |
Lead Time 测试周期: 12 working days
Sample size 送样规格: 70pcs
FDA更新抽样要求
It has recently come to our attention that 5 samples for Bacterial Filtration Efficiency (BFE), Differential Pressure (Delta P), and Particle Filtration Efficiency (PFE) may no longer be adequate for FDA submissions. In the past, many manufacturers tested a minimum of 5 samples for Delta P, BFE, and PFE and historically this amount has been successful in getting submissions approved through FDA. But with recent information we have received from FDA and multiple sponsors, is it clear that FDA is enforcing requirements for sampling as stated in the ASTM F 2100 standard as listed below:
7. Sampling
7.1 Testing shall be performed on materials taken from manufactured face masks.
7.2 An acceptable quality limit of 4% shall be used for all required testing to establish conformance of medical face masks to a specific performance class.
7.3 Examples of acceptable sampling plans are found in ANSI/ASQC Z1.4 and ISO2859-1.
8. Number of Tests
8.1 A sufficient number of medical face mask shall be evaluated for each test to achieve the established acceptable quality limit or confidence level.
Beyond the ASTM F2100 specification, FDA has also been asking manufactures to test up to three non-consecutive lots to demonstrate lot-to-lot performance or justify why this was not performed. For FDA submissions, we recommend you take the most conservative route to prevent any delays in your submission. The table below is our current recommendation based on recent trends:
Test | Standard | Sample amount per lot | # of lots | Total |
Synthetic Blood | ASTM F1862 (specifies 32) | 32 | 3 | 96 |
Flammability | 16 CFR Part 1610 (specifies 14) | 14 (4 preliminary, 5 preliminary classification, 5 if necessary, based on burn time) FDA is requiring 32 samples | 3 | 100 |
BFE | ASTM F2101 (no sample amount specified) | AQL 4%, FDA will accept 32 | 3 | 96 |
Delta P | EN 14683 (no sample amount specified) | AQL 4% FDA will accept 32 | 3 | 96 |
PFE | ASTM F2299 (no sample amount specified) | AQL 4% FDA will accept 32 | 3 | 96 |
Five samples may be sufficient for periodic R&D or verification testing but is not recommended for FDA submission.